Featuring all the advantages of the HPC 40, this pressure calibrator takes it up a notch and offers better battery life, improved display, and an external module connection, all while being fully compensated for pressure, temperature, and electrical measurements at -4 to 122°F (-20 to 50°C). Carrying case not included with "no pump" option.
Featuring all the advantages of the HPC 40, this pressure calibrator takes it up a notch and offers better battery life, improved display, and an external module connection, all while being fully compensated for pressure, temperature, and electrical measurements at -4 to 122°F (-20 to 50°C). Carrying case not included with "no pump" option.
With intrinsic safety as top priority, this dual handheld pressure calibrator comes with options to add an additional two external pressure or temperature modules.
Features
Applications
psi (Gauge Pressure) 18 to 28°C | |
18 to 28°C | 0 to 30% of range ±(0.01% of full scale) |
30 to 110% of Range | ±(0.035% of reading) |
Vacuum | ±(0.05% of full scale) |
psi (Gauge Pressure) -20 to 50°C | |
0 to 30% of Range | ±(0.015% of full scale) |
Vacuum | 30 to 110% of range |
psi (Absolute Pressure with BARO Option) | |
15 psi Range | Gauge accuracy +0.005 psiA |
30 psi Range | Gauge accuracy +0.005 psiA |
100 psi Range | Gauge accuracy +0.002 psiA |
Differential Pressure | |
Without Tare Function | ±(0.05% of static line pressure reading) |
Pressure Sensor | |
Wetted Materials | 316 stainless steel (wrench tight) 316 stainless steel (finger tight) and Viton® with internal O-ring (10 psi/1 bar/100 kPa) 316 stainless steel and Viton® |
Diaphragm Seal Fluid | Silicone oil |
Connection | Crystal CPF female |
Barometric Reference (Baro) | |
Accuracy | ±0.00725 psi, ±0.5 mbar |
Range | 10.153 to 15.954 psiA 700.0 to 1100.0 mbarA |
Units and Resolution | psi: 0.001 inHg: 0.001 mmHg: 0.01 mbar: 0.1 |
Current (mA) Input | |
Accuracy | ±(0.015% of rdg + 0.002 mA) |
mA Range | 0 to 55 mA |
Percent Range | 0 to 20, 4 to 20, 10 to 50 |
Maximum Allowable Current | 93.3 mA |
Resolution | 0.001 mA or 0.01% |
Units | mA, scaling, % error, and % flow |
Input Resistance | <4.99 Ω |
Voltage Burden at 20 mA | <0.10 V |
Voltage Burden at 50mA | <0.250 V |
HART Resistance | 250 Ω |
Current (mA) Sink | |
Accuracy | ±(0.015 of rdg + 0.002 mA) |
Range | 0 to 25 mA |
Step Time | 1 to 999 seconds |
Ramp Time | 5 to 999 seconds |
Voltage (V DC) Input | |
Accuracy | ±(0.015 % of rdg + 2 mV) |
Range | 0 to 28 V DC |
Resolution | 0.001 V DC |
Switch Test | |
Switch Type | Dry contact |
Closed State Resistance | <1K Ω |
Open State Resistance | >100K Ω |
Sample Rate | 10 Hz |
General Specifications | |
Temperature | Operating: -4 to 122°F (-20 to 50°C) Storage: -40 to 167°F (-40 to 75°C) |
Power | 1.5 V alkaline batteries (cell voltage) |
Battery Life | No external modules: 35 hours One external module: 25 hours Two external modules: 12 hours |
Housing | PC/PBT plastic |
Rating | IP66/67 |
Weight | 1.25 lbs (567 g) |
psi (Gauge Pressure) 18 to 28°C | |
18 to 28°C | 0 to 30% of range ±(0.01% of full scale) |
30 to 110% of Range | ±(0.035% of reading) |
Vacuum | ±(0.05% of full scale) |
psi (Gauge Pressure) -20 to 50°C | |
0 to 30% of Range | ±(0.015% of full scale) |
Vacuum | 30 to 110% of range |
psi (Absolute Pressure with BARO Option) | |
15 psi Range | Gauge accuracy +0.005 psiA |
30 psi Range | Gauge accuracy +0.005 psiA |
100 psi Range | Gauge accuracy +0.002 psiA |
Differential Pressure | |
Without Tare Function | ±(0.05% of static line pressure reading) |
Pressure Sensor | |
Wetted Materials | 316 stainless steel (wrench tight) 316 stainless steel (finger tight) and Viton® with internal O-ring (10 psi/1 bar/100 kPa) 316 stainless steel and Viton® |
Diaphragm Seal Fluid | Silicone oil |
Connection | Crystal CPF female |
Barometric Reference (Baro) | |
Accuracy | ±0.00725 psi, ±0.5 mbar |
Range | 10.153 to 15.954 psiA 700.0 to 1100.0 mbarA |
Units and Resolution | psi: 0.001 inHg: 0.001 mmHg: 0.01 mbar: 0.1 |
Current (mA) Input | |
Accuracy | ±(0.015% of rdg + 0.002 mA) |
mA Range | 0 to 55 mA |
Percent Range | 0 to 20, 4 to 20, 10 to 50 |
Maximum Allowable Current | 93.3 mA |
Resolution | 0.001 mA or 0.01% |
Units | mA, scaling, % error, and % flow |
Input Resistance | <4.99 Ω |
Voltage Burden at 20 mA | <0.10 V |
Voltage Burden at 50mA | <0.250 V |
HART Resistance | 250 Ω |
Current (mA) Sink | |
Accuracy | ±(0.015 of rdg + 0.002 mA) |
Range | 0 to 25 mA |
Step Time | 1 to 999 seconds |
Ramp Time | 5 to 999 seconds |
Voltage (V DC) Input | |
Accuracy | ±(0.015 % of rdg + 2 mV) |
Range | 0 to 28 V DC |
Resolution | 0.001 V DC |
Switch Test | |
Switch Type | Dry contact |
Closed State Resistance | <1K Ω |
Open State Resistance | >100K Ω |
Sample Rate | 10 Hz |
General Specifications | |
Temperature | Operating: -4 to 122°F (-20 to 50°C) Storage: -40 to 167°F (-40 to 75°C) |
Power | 1.5 V alkaline batteries (cell voltage) |
Battery Life | No external modules: 35 hours One external module: 25 hours Two external modules: 12 hours |
Housing | PC/PBT plastic |
Rating | IP66/67 |
Weight | 1.25 lbs (567 g) |